ID: 9098075
UV-VIS Spectrophotometer Wavelength range: 190 - 1100 nm Slit width: 1 nm EP Resolution test: > 1.6 toluene in hexane Ratio ABS: 269 nm/ 266 nm Stray light: NaNO2, ASTM: <0.03% at 340 nm Nal, ASTM: <0.05% at 220 nm KCl, EP: <1% at 198 nm Wavelength accuracy: <±0.5 nm at 0.5/sec scan (NIST 2034) <±0.2 nm at 486.0 and 656.1 nm Wavelength reproducibility: < ±0.02 nm Photometric accuracy: < ±0.005 A at 440.0, 465.0, 546.1, 590.0, 635.0 nm, 1 A (NIST 930e) Potassium dichromate and EP method: < ±0.01 A at 235, 257, 313, 350, 430 nm Photometric noise: <0.0002 A, 60.5/sec scans at 0 A, 500 nm/rms Photometric stability: <0.001 A/h at 0 A, 340 nm Baseline flatness: < 0.001 A Scan time: 1-5/sec (full range) Computer Monitor Operating system: Windows 7 Pro Manuals Cables.
AGILENT/HP/HEWLETT-PACKARD/KEYSIGHT 8453 분광사진계는 스펙트럼 분석을위한 강력하고 정확한 도구입니다. 그것은 자외선 및 가시광선 스펙트럼에서 전자기 방사선의 흡광도와 투과율을 모두 측정함으로써 작동합니다. HP 8453의 파장 범위는 190 ~ 1100nm이며, 샘플을 통과하는 광자를 정확하게 측정하고, 이러한 판독값을 흡광도 및 투과율 데이터로 변환할 수 있습니다. AGILENT 8453은 샘플 측정의 정확성과 정확성을 보장하는 기기를 사용하기 쉽습니다. 보정 된 흰색 광원과 차폐, 냉각, 안정화 된 검출기 시스템을 사용하여 빛을 수집하고 분석합니다. 그런 다음, 이 빛 이 "샘플 '을 통과 하게 되고," 샘플' 을 통과 한 후 에 흡수 되어 전달 되는 빛 의 양 이 결정 된다. 그런 다음 정보는 흡광도 또는 투과율 그래프 대 파장으로 변환됩니다. 8453에는 매우 다재다능한 도구로 만드는 고급 기능도 있습니다. 여기에는 자동 교정, 파장 스캔, 온보드 진단 및 프로그래밍 가능한 파장 측정이 포함됩니다. 이러한 기능은 사용자가 정확성, 재현성 및 유연성을 높일 수 있도록 해줍니다. 또한 사용자 편견을 제거하고 안정적인 결과를 제공하는 GLP/GMP/ISO 규정도 준수합니다. HEWLETT-PACKARD 8453 (HEWLETT-PACKARD 8453) 은 비용 효율적인 패키지에서 우수한 성능을 필요로 하는 실험실 및 기타 연구 시설에 적합한 선택입니다. 고급 기능, 정확성 및 정확성을 갖춘 KEYSIGHT 8453은 환경 모니터링, 식품 품질, 안전, 제약 테스트 등의 분야에서 다양한 응용 분야에 적합합니다.
아직 리뷰가 없습니다